Page 241 - Mahr马尔量具完整
P. 241

General Technical Data of Inductive Probes


           The measuring principle of inductive probes is based   The Mahr P2000 series of probes applies a high linear,
          on the change of position of the magnet‘s conductive
                                                               patented VLDT transducer which is similar to an LVDT
          core moving within a coil system. Generally, this is   transducer. This also operates according to a differential
          distinguished between a half-bridge and LVDT‘s.      transformer principle.








                                  S                       S                        S







                               Half-bridge HB              LVDT                     VLDT
                           (Differential Choke Coil)   (Linear Variable           (Very Linear
                                                    Differential Transducer)  Differential Transducer)



         Electrical specification of various compatibilities


                                         Type           Mahr            Tesa           Marposs      Mahr-Federal
          Carrier frequency  KHz                         19.4            13               7.5             5
                                         P2001
                                         P2004          192              73.75          115              78.74
                                         P2104
                                         P1300          192              73.75             —               —
          Sensitivity        mV / V / mm  1301
                                         1303           192                —               —               —
                                         1304 K
                                         1318
                                         P2010           19.2            29.5            11.5             7.874
          Amplitude          V eff                        5               3               3.5             2


        Schematic drawings of Mahr input amplifiers according to the various
        compatibilities



                                                        Connector pin
                                                        (contact side)
                                    Oscillator    Mahr            Oscillator    Tesa









                                                  Marposs                       Mahr-Federal
                                    Oscillator                    Oscillator












                                                       Millimar | Electrical Length Measuring Instruments      237
   236   237   238   239   240   241   242   243   244   245   246