Page 10 - Mahr马尔量具完整
P. 10
MarCal | Terms
Measuring blades for
inside measurement Locking screw Depth bar
Slide Beam
Vernier scale
Line scale
Movable measuring jaw
Fixed
measuring
jaw
Integrated
wireless data High contrast analog display Rounded
Digit size: 11 mm depth bar
Intuitive handling
by menu guidance
Switch on: Press the ON key
Lock function or move the slide
(key lock) RESET (Set display to zero)
MarCal | Types of measurement Error limits G according to DIN 862
16 series calipers (e.g. 16 EWRi, 16 U, 16 FN) can be used Error limits G in µm
to conduct 4 different types of measurements:
Measuring Scale interval and/or Resolution
length vernier interval
a) Outside
measurement mm 0.1 and 0.05 0.02 0.01
50
20 20
100
50
200
b) Inside
measurement 300
400 60 30 30
500 70
600 80
c) Depth 700 90
measurement 800 100
40 40
900 110
1000 120
1200 140 50
d) Step (Distance) 1400 160
measurement
1600 180 —
60
1800 200
2000 220
6 MarCal | Caliper